od roku 1992 Vám prinášame "riešenia na presné meranie."
  • sk
Kategórie

Fluke PM6304 - RCL Meter

Fluke PM6304 - RCL Meter

Cena: na požiadanie


Dostupnosť: na požiadanie


Poslať známemu


PM 6304 RCL Meters
Versatile component measurement and testing

If you want an instrument with remote programmability, consider the Fluke PM 6304 RCL meter. The PM 6304 has a basic accuracy of 0.1%, 3 test voltage levels and 204 test frequencies. The PM 6304C has an even higher basic accuracy, and can measure components or devices with an accuracy of 0.05%. Thanks to this high accuracy, the PM 6304C is idealy suited for calibrating the capacitance sourcing function of the high-performance Fluke 5500A Multi-Product Calibrator.

PM 6304 Programmable RCL meter (DC-100 kHz)

AC Test Mode
Test frequency
PM 6306:50, 60, 100, 120 Hz
 200 Hz to 100 kHz (100 Hz steps)
 100 kHz to 1 MHz (1 kHz steps)
PM 6304:50, 60, 100, 120 Hz
 200 Hz to 20 kHz (100 Hz steps)
 100 kHz
Test frequency accuracy
PM 6306:0.01%
PM 6304:0.01%
Test signal levels
PM 6306:50 mV to 2V (10 mV steps) via 100Ω
PM 6304:50 mV via 100Ω
 1V via 100Ω
 2V via 400Ω
Basic measurement accuracy at normal measurement mode
PM 6306:0.1% ± 1 digit (for ³ 0.25V, £ 50 kHz)
 0.1% * (f/50 kHz) ± 1 digit (for ³ 0.25V, > 50 kHz)
 0.1% * (0.25V/Vt) ± 1 digit (for < 0.25V, £ 50 kHz)
PM 6304:0.05% ± 1 digit (for PM 6304C, £ 2 kHz)
 0.1% ± 1 digit (for £ 20 kHz)
 0.4% ± 1 digit (100 kHz)
 0.5% ± 1 digit (for 50 mV £ 20 kHz)
 2.0% ± 1 digit (for 50 mV, 100 kHz)
DC bias
Internal
PM 6306:0 to 10V (0.1V steps)
PM 6304:2V
External
PM 6306:0 to 40V
PM 6304:0 to 40V
DC test mode (Optional)
Test signal levels
PM 6306:50 mV to 2V (10 mV steps) via 100Ω
PM 6304:300 mV via 100Ω
 1V via 100Ω
 2V via 400Ω
Basic measurement accuracy
PM 6306:0.1% ± 1 digit (for ³ 0.25V)
PM 6304:0.1% ± 1 digit (for 1V)
Contact check (PM 6306 only)
Pass
 3 3Ω
Fail
 3Ω (with indication of failed connection lead)
Maximum measuring ranges
Impedance/Resistance AC Z or RAC
 0.0000Ω to 200 MΩ
Resistance DC R DC
 0.0000Ω to 50 MΩ
Capacitance C
 0.00 pF to 31.8F
Inductance L
 0.00 mH to 637 kH
Quality factor Q
 0.000 to 1000
Dissipation factor D
 0.000 to 1000
Phase Angle F
 -179 to + 180 deg
Voltage monitor VX
 0.1 mV to 2.00V
Current monitor IX
 0.005 mA to 10.0 mA
Maximum resolution
Impedance/Resistance AC Z or RAC
 0.1 mΩ
Resistance DC R DC
 0.1 mΩ
Capacitance C
 0.01 pF
Inductance L
 0.01 mH
Quality factor Q
 0.001
Dissipation factor D
 0.001
Phase Angle F
 0.1 deg
Voltage monitor VX
 0.1 mV
Current monitor IX
 0.001 mA
Circuit diagram
Display
 1 of 7 different equivalent circuit diagrams
Auto Mode
Read-out
 Dominant and secondary parameter
Equivalent circuit diagram
 Parallel for R+C, Serial for R+L
Manual mode
Read-out
 Dominant and secondary parameter or Z, F, D, Q, VX, IX
Equivalent circuit diagram
 Parallel or serial selectable
Average function
Function
 Exponential averaging in continuous mode
Levels
PM 6306:3 (and off)
PM 6304:1 (and off)
Deviation mode (PM 6306 only)
Relative range in respect to reference value
 -100% to +100%
Measuring modes (Normal)
Continuous
 2 measurements/s
Single
 Triggered via "TRIG" key
 Triggered via handler interface
 Triggered via IEEE-488 or RS-232
Test frequency
PM 6306:50, 60, 100, 120 Hz
 200 Hz to 100 kHz (100 Hz steps)
 100 kHz to 1 MHz (1 kHz steps)
 DC (optional)
PM 6304:50, 60, 100, 120 Hz
 200 Hz to 20 kHz (100 Hz steps)
 100 kHz
 DC (optional)
Read-out
 Display or via IEEE-488 or RS-232 interface
Measuring Modes (Fast)
Max. speed
 10 measurements/s
Test frequency
PM 6306:200 Hz to 100 kHz (200 Hz steps)
 100 kHz to 1 MHz (1 kHz steps)
 DC (optional)
PM 6304:200 Hz to 20 kHz (200 Hz steps) 100 kHz
 DC (optional)
Single
 Triggered via handler interface
 Triggered via IEEE-488 or RS-232
Read-out
 Via IEEE or RS-232 interface (display blanked)
Binning
Standard bins
 9
Special bins
 Bin "0" and bin "fail"
Bin programming via
 IEEE-488 interface
 RS-232 interface
 Bin programmer (PM 6304 only)
Bin limit programming
 Absolute or relative
Trim function
Open circuit
 Open circuit compensation Z > 100 kW
Short circuit
 Short circuit compensation Z < 10W
Protection against charged capacitors
C < 2 mF
PM 6306:Vmax < 200V
PM 6304:Vmax < 500V
2 mF o C o 2 mF
PM 6306:Vmax <47 x (C/mF) -0.234
 (Vmax in V, C in mF)
PM 6304:Vmax <117 x (C/mF) -0.234
 (Vmax in V, C in mF)
Stored settings (non-volatile memory)
Front panel settings
 9 + 1 (trim figures included)
Bin settings
 9 + 1
Print measurements results
Print measurements results
 Via RS-232 interface for serial printers
Calibration interval
Calibration interval
 1 year
Environmental Conditions
Operating Temperature
 0° to 50° C
Storage Temperature
 -40° C to 70° C
Power, battery life
 100/120/220/240V + 10%
Line frequency
 50/60 Hz
Power Consumption
PM 6306:44 VA
PM 6304:31 VA
Safety
 According to CE-regulation 73/23 EN61010-1 CAT II, Pollution Degree 2
EMC
 According to CE regulation 89/336: Emission according to EN 55011, Group 1 Class B, respectively CISPR 11. Immunity according to EN 50082-1, inclusive IEC 801-2, -3, -4
Warm up Time
PM 6306:30 minutes
PM 6304:5 minutes
Dimensions and weight
W x H x D
 315 x 105 x 405 mm (12.4 in x 4.13 in x 15.9 in)
Weight
PM 6306:5.3 kg (11.7 lb)
PM 6304:4.7 kg (10.4 lb)




- RCL  -  

30 produktov v rovnakej kategórii:

Novinky a zmeny

séria 3300W zdrojov SM3300


najlepšie parametre výstupného prúdu