The INCU™ is an incubator analyzer, designed around AAMI and IEC standards, that specifies temperature, humidity, sound and airflow characteristics of incubators.

  • Portability (sitting in place of infant within incubator)
  • Single-instrument measurement and documentation for key infant-incubator parameters
  • Measurement for most parameters according to IEC and AAMI standards
  • Time-saving in testing critical infant incubators
  • Windows® software for evaluation and documentation
  • Numerical and graphical reports
  • Adjustable measurement intervals
  • Standalone, or full function (computer required)
INCU™ Incubator Analyzer
Specifications
Display
  2 x 16 super twist LCD
Resolution
  1470 measurement points (210 records of seven parameters)
Power Supply
  Lead-acid battery for 24 hours continuous operation
  Battery charger 600 mA min, AC-DC 6 V
Sound Level
  30 to 80 dBA
Relative Humidity [without condensation]
Measuring Range: 0 % RH to 100 % RH
Resolution: 0.1 % RH
Accuracy: 0 % to 90 % ± 5 % RH, 25 ºC to 40 ºC ± 5.3 %
Four Temperatures
  5 ºC to 70 ºC
Conduction - one sensor in contact with mattress
Convection (inside the incubator) - three sensors (T1, T3, T4)
Convection (inside or outside the incubator) - one sensor (T2)
Radiation - one sensor for radiation head that can be used also for external temperature
Airflow
  0.1 to 0.7 m/s
Temperature: reading value ± 0.5 ºC ± 1 LSB for the range from 25 ºC to 40 ºC
Airflow: from 0.1 m/s to 0.5 m/s reading ± 0.1 m/s at temperature 25 ºC to 40 ºC and humidity 50 % RH ± 15 % RH
Dimensions
  10.6 in L x 7.8 in W x 5.5 in H
  (27 cm L x 20 cm W x 14 cm H)
Weight
  6.6 lb
  (3 kg)
INCU software PC system requirements
  • Pentium 133 MHz microprocessor
  • 16 MB free hard disk space
  • SVGA color monitor
  • Serial port or USB port (may require serial to USB adapter, ICON USB 2.0 to RS-232 cable recommended (sourced separately)
  • Mouse compatible with Microsoft® operating system
  • 1 CD ROM disk drive
  • Microsoft Windows 95B, Windows 98, Windows NT 4.0, or Windows 2000
  • Laser or inkjet printer