od roku 1992 Vám prinášame "riešenia na presné meranie."
  • sk
Kategórie

Fluke PM6303 - RCL Meter

Fluke PM6303 - RCL Meter

Cena: na požiadanie


Dostupnosť: na požiadanie


Poslať známemu


RCL Meter

Effective, economic testing that's quick and easy to use

The PM 6303A is quick and simple to operate, on the manufacturing floor as a quality assurance tool or in the service lab for quick tests of a wide variety of components. Its simple operation and 0.25% basic accuracy make the PM 6303A equally at home in the classroom for education and training environments.

PM 6303A/01 RCL Meter (1kHz)

Technical Specifications
AC test mode
Test frequency:1 kHz
Test frequency accuracy:0.025%
Test signal level:2V via 400Ω source
Basic measurement accuracy:0.25% ± 1 digit
DC bias
Internal:2V
Maximum measurement ranges
Impedance/Resistance Z or RAC:0.000Ω to 200 MΩ
Capacitance C:0.0 pF to 100 mF
Inductance L:0.0 mH to 32 kH
Quality factor Q:: 0.0002 to 500
Dissipation factor D:0.002 to 500
Phase angle F:-90.0 to +90.0 deg
Maximum resolution
Impedance/Resistance Z or RAC:01 MΩ
Capacitance C:0.1 pF
Inductance L:0.1 mH
Quality factor Q:0.001
Dissipation factor D:0.001
Phase angle F:0.1 deg
Circuit diagram
 7 different equivalent circuit diagrams
Auto mode
Read-out:Dominant parameter
Equivalent circuit diagram:Parallel for R+C
 Serial for R+L
Manual Mode
Read-out:Dominant or secondary parameter
Equivalent circuit diagram:Parallel or serial selectable
Measurement update rate
 2 measurements/s
Trim function
Open circuit:Open circuit compensation
 Z > 100 kΩ
Short circuit:Short circuit compensation
 Z < 10Ω
Stored settings (non-volatile memory)
Front panel settings:1 (trim figures included)
Calibration Interval
 1 year
Environmental Conditions
Operating temperature:0°C to 50°C
Storage temperature:-40°C to 70°C
Power requirements:100/120/220/240V ± 10%
Line frequency:50 to 100 Hz
Power consumption:16 VA
Safety:According to CE-regulation 73/23 EN61010-1 CAT II, Pollution Degree 2
EMC:According to CE regulation 89/336: Emission according to EN 55011, Group 1 Class B, respectively CISPR 11. Immunity according to EN 50082-1, inclusive IEC 801-2, -3, -4
Warm-up time:5 minutes
Dimensions and weight
W x H x D:315 x 105 x 401 mm (12.4 in x 4.13 in x 15.9 in)
Weight:3.8 kg (8.4 lb)




- RCL  -  

30 produktov v rovnakej kategórii:

Novinky a zmeny

séria 3300W zdrojov SM3300