If you need a precision measurement at low temperatures, do not look any further than Hart Scientific.
The 5623B, precision “freezer probe,” is specially sealed from the sensing element to the end of the probe cable, preventing ingress of moisture when exposed to temperatures as low as –100 °C. The entire assembly withstands temperatures over its full range (–100 °C to 156 °C), which is ideal for verification of freezers or autoclaves where a thermo-well isn’t available. The 5623B assembly can be fully immersed in fluids when the application may require use in a liquid bath. The 5623B is available in a 6.35 mm (0.25 in) dia. × 125 mm (6 in) long Inconel™ sheath. With accuracy (that includes calibration uncertainty and short-term drift) of ± 0.05 °C over its full range, the 5623B is just right as a secondary standard for calibration of other process sensors.
Most Hart Scientific readouts make an excellent companion for the 5623B. We recommend the use of the 1521, 1522, 1502A, 1529, or 1560 thermometer readouts.
With each 5623B, you receive a full NVLAP-accredited calibration report. This report includes test data and ITS-90 calibration coefficients to enter into your Hart Scientific thermometer readout.
|
Specifications |
|
|
Resistance |
Nominal 100 W (± 0.1 W) |
|
Temperature Coefficient |
0.003925 ohms/ohm/ °C nominal |
|
Temperature Range |
–100 °C to 156 °C |
|
Transition Temperature |
–100 °C to 156 °C |
|
Drift Rate |
± 0.01 °C per year maximum at 0 °C, when used periodically at max temperature |
|
Sheath Material |
Inconel™ 600 |
|
Leads |
Teflon™-insulated, silver-plated stranded copper, 22 AWG. |
|
Termination |
Specify. See ordering information. |
|
Calibration |
Includes manufacturer’s NVLAP-accredited calibration and table with R vs. T values in 1 °C increments from –80 °C to 156 °C. ITS-90 coefficients included. |
|
Probe Accuracy (includes calibration uncertainty and short-term stability) |
± 0.05 °C over the full range |
|
Cable Length |
6.7 meters (20 ft) |
|
Size |
6.35 mm (0.25 in) dia. x 152 mm (6 in) |
|
Ordering Information |
|
|
5623B-6-X |
Freezer Probe, RTD 1/4" dia. x 6", –100 to 156°C |
|
2601 |
Probe Carrying Case |
|
X = termination. Specify “B” (bare wire), “D” (5-pin DIN for Tweener Thermometers), |
|
|
|






























teplota
teplomer
zdroj
vlhkomer
vlhkosť
USB
osciloskop
multimeter
biocote
Automotive
antibakteriálny
záznamník
pyrometer
vodotesný
EN13485
IP65
robustný
IP68
ponorný
cukrársky
generátor
2chan
PC
AWG
True RMS
logger
scopemeter
DC
infračervený
rosný bod
software
Fluke
datalogger
4-20mA
kontaktný
HACCP
kalibrátor
zadávač
sonda
js
wireless
transmitter
IR
4Chan
pleseň
termokamera
vpichový
12bit
dataloger
tlak
metrologický
200MHz
dozimeter
modulácia
dmm
IEEE 488
GPIB
pH
laboratórny
100MHz
potraviny
SCC Kit
termočlánok
komunikácia
IR-Fusion
kvalita siete
softvér
referenčný
pasterizácia
čas
EN50160
presný
LAN
digitálny
odmerka
váha
250MHz
PQ
SW
objektív
analyzátor
PS Cumulus Flash tag cloud adapted from the Wordpress original by Roy Tanck requires Flash Player 9 or better. Ported by ecartservice.net.