If cuteness were reason enough to buy a product, Hart’s Mini Fixed-Point Cells would win you over easily. But there’s a much better reason to buy them: they give you the least expensive, easiest-to-use fixed-point standards for your lab.
Mini cells eliminate the need for comparison calibrations. Temperatures of fixed-point cells are constant and intrinsic, so only the electrical parameters of the sensor under calibration need to be read. If you’re calibrating industrial thermometers, thermocouples, or thermistors and want the most accurate calibration possible, these mini cells will give it to you. If you need a wide range of temperatures, mini cells cover the triple point of water (0.01 °C) and every ITS-90 point from indium (156.5985 °C) to copper (1084.62 °C).
With mini cells, realization and maintenance are simple. Mini TPW cells can be automatically realized and maintained in our 9210 Maintenance Apparatus (page 30). Realizing the triple point of water takes only five minutes, but the plateaus last all day.
The realization and maintenance of indium, tin, zinc, and aluminum cells are likewise automated through our 9260 Mini Fixed-Point Cell Furnace (page 32). Work with them at their designated freeze point, or use them at their melting point to simplify the calibration process even further. We published a paper, “The Comparison Between the Freezing Point and Melting Point of Tin,” to help you understand and benefit from the easier procedure of using the melting point of your standard.
These mini cells are made from the same materials and with the same procedures as their full-size counterparts. In fact, they can achieve nearly the same uncertainty levels as Hart’s traditional fixed-point cells. Probes as short as nine inches work with these cells. The specifications table (at right) gives you the immersion depth and uncertainty for each cell.
In addition to high-accuracy calibrations of RTDs and PRTs, these cells are perfect for validating the accuracy of SPRTs. If you’re doing comparison calibrations with SPRTs, then you know the importance of occasionally checking their accuracy between their own recalibrations. Because these cells are easy to use and maintain, verification checks are simple and convenient.
Metal-cased cells can also be used in the 9260 maintenance furnace. Because they use stainless steel cases, these cells are easier to use and transport without risk of breakage. You’ll notice that we have designed the metal cased cells with more immersion depth to give even better uncertainty too!
You’ll find these cells easier to use than you expect. You can have a free copy of Xumo Li’s paper comparing freeze-point measurements with melting-point measurements, and if you want a high level of training in using metal freeze-point cells, you can attend one of Hart’s in-depth training classes held in our lab in Utah.
Specifications |
||||||||
|
|
Uncertainty (mK) k=2 |
|||||||
|
|
|
|
|
Inside Diameter |
Total Cell Height |
Immersion Depth1 |
|
Simple |
|
5901B-G |
Water |
0.01 |
30 mm |
8 mm |
170 mm |
117 mm |
0.2 |
0.5 |
|
5914A |
Indium |
156.5985 |
43 mm |
8 mm |
214 mm |
140 mm |
1.0 |
2.0 |
|
5915A |
Tin |
231.928 |
43 mm |
8 mm |
214 mm |
140 mm |
1.4 |
3.0 |
|
5916A |
Zinc |
419.527 |
43 mm |
8 mm |
214 mm |
140 mm |
1.6 |
4.0 |
|
5917A |
Aluminum |
660.323 |
43 mm |
8 mm |
214 mm |
140 mm |
4.0 |
10.0 |
|
5918A |
Silver |
961.78 |
43 mm |
8 mm |
214 mm |
140 mm |
7.0 |
n/a |
|
5919A |
Copper |
1084.62 |
43 mm |
8 mm |
214 mm |
140 mm |
15.0 |
n/a |
|
5944 |
Indium |
156.5985 |
41.3 mm |
7.8 mm |
222 mm |
156 mm |
0.7 |
1.4 |
|
5945 |
Tin |
231.928 |
41.3 mm |
7.8 mm |
222 mm |
156 mm |
0.8 |
1.6 |
|
5946 |
Zinc |
419.527 |
41.3 mm |
7.8 mm |
222 mm |
156 mm |
1.0 |
2.0 |
| 5947 | Aluminum F. P. |
660.323 | 41.3 mm | 7.8 mm | 222 mm | 156 mm | 2.0 | 4.0 |
| 1Distance from the bottom of the central well to the surface of the pure metal.
2“Cell Only” refers to the expanded uncertainty of the cell when realized by traditional methods and maintained using traditional maintenance devices. “Simple Realization” refers to the expanded uncertainty of the cell when realized using practical methods (melting points instead of freezing points or slush ice instead of an ice mantle, for example) and maintained using Hart’s 9210 and 9260 mini cell maintenance apparatus. |
||||||||
|
||||||||||||||||||||||||||||
|
|






























teplota
teplomer
zdroj
vlhkomer
vlhkosť
USB
osciloskop
multimeter
biocote
Automotive
antibakteriálny
záznamník
pyrometer
vodotesný
EN13485
IP65
robustný
IP68
ponorný
cukrársky
generátor
2chan
PC
AWG
True RMS
logger
scopemeter
DC
infračervený
rosný bod
software
Fluke
datalogger
4-20mA
kontaktný
HACCP
kalibrátor
zadávač
sonda
js
wireless
transmitter
IR
4Chan
pleseň
termokamera
vpichový
12bit
dataloger
tlak
metrologický
200MHz
dozimeter
modulácia
dmm
IEEE 488
GPIB
pH
laboratórny
100MHz
potraviny
SCC Kit
termočlánok
komunikácia
IR-Fusion
kvalita siete
softvér
referenčný
pasterizácia
čas
EN50160
presný
LAN
digitálny
odmerka
váha
250MHz
PQ
SW
objektív
analyzátor
PS Cumulus Flash tag cloud adapted from the Wordpress original by Roy Tanck requires Flash Player 9 or better. Ported by ecartservice.net.