The gallium melting point (29.7646 °C) is a critical temperature. Thermometers used in life science, environmental monitoring, and many other applications depend on it for accurate calibrations. Lab standards rely on it as an ITS-90 check standard and as a means of measuring drift between calibrations. Hart Scientific now makes it easy to use.
The new 9230 Gallium Maintenance System works with Hart’s Model 5943 Stainless Steel Gallium Cell to provide melting plateaus that last a week, with results approaching what can be achieved in a Hart maintnance bath. Not a day. Not a day-and-a-half. One week.
The Model 5943 Stainless Steel Gallium Cell holds a gallium sample that is 99.99999+ % pure. The gallium is sealed in a Teflon envelope in a high purity argon atmosphere, which is itself sealed inside a stainless steel housing. This double-sealing method reduces leaching into the gallium sample and ensures a life of ten years or longer for the cell.
Specifications |
|
|
Temperature Range |
15 °C to 35 °C |
|
Ambient Operating Range |
18 °C to 28 °C |
|
Stability |
±0.02 °C |
|
Vertical Gradient |
< 0.03 °C over six inches during cell maintenance |
|
Plateau Duration |
Five days, typical |
|
Resolution |
0.01 ° (0.001 ° in program mode) |
|
Display Scale |
°C or °F, switchable |
|
Immersion Depth |
220 mm (8.75 in) in gallium cell |
|
Stabilization Time |
Preprogrammed |
|
Preheat Wells |
2 |
|
Fault Protection |
Heating/cooling rate cutout |
|
Display Accuracy |
±0.05 °C at 29.76 °C |
|
Comparison Block |
Contact Hart |
|
Well-to-Well Gradient (in comparison block) |
n/a |
|
Heating Time |
Preprogrammed |
|
Cooling Time |
Preprogrammed |
|
Comm. |
RS-232 included |
|
Power Requirements |
115 VAC (±10 %), 60 Hz, 1.0 A, or |
|
Exterior Dimensions |
489 x 222 x 260 mm |
|
Weight |
8.2 kg (18 lb.) without cell |
|
||||||||
|
|






























teplota
teplomer
zdroj
vlhkomer
vlhkosť
USB
osciloskop
multimeter
biocote
Automotive
antibakteriálny
záznamník
pyrometer
vodotesný
EN13485
IP65
robustný
IP68
ponorný
cukrársky
generátor
2chan
PC
AWG
True RMS
logger
scopemeter
DC
infračervený
rosný bod
software
Fluke
datalogger
4-20mA
kontaktný
HACCP
kalibrátor
zadávač
sonda
js
wireless
transmitter
IR
4Chan
pleseň
termokamera
vpichový
12bit
dataloger
tlak
metrologický
200MHz
dozimeter
modulácia
dmm
IEEE 488
GPIB
pH
laboratórny
100MHz
potraviny
SCC Kit
termočlánok
komunikácia
IR-Fusion
kvalita siete
softvér
referenčný
pasterizácia
čas
EN50160
presný
LAN
digitálny
odmerka
váha
250MHz
PQ
SW
objektív
analyzátor
PS Cumulus Flash tag cloud adapted from the Wordpress original by Roy Tanck requires Flash Player 9 or better. Ported by ecartservice.net.