Tento prístroj nespĺňa RoHs a v Európe sa nedodáva, bol nahradený modelom Fluke 8588A
.
.
.
Hewlett Packard 3458A
Keysight Technologies 3458A
Agilent 3458A - HFL
Fluke HP 3458A - HFL
multimeter HP 3458
.
Technical Specifications |
Voltage DC |
|
Resistance |
Range: |
0 to 1.2 GΩ |
Best 1 Year Specification: |
±7.5 ppm of reading |
|
Current DC |
Range: |
0 to ±1.05A |
Best 1 Year Specification: |
±20 ppm of reading |
|
Voltage AC |
Range: |
1 mV to 700V |
|
1 Hz to 10 MHz, Sine |
Best 1 Year Specification: |
±0.007% of reading |
|
Current AC |
Range: |
1 mV to 700V |
|
1 Hz to 10 MHz, Sine |
Best 1 Year Specification: |
±0.007% of reading |
|
General Math Function Specifications |
Null |
X-OFFSET |
|
Minimum Execution Time = |
180 µs |
|
PERC |
100 x (X-PERC)/PERC |
|
Minimum Execution Time = |
600 µs |
|
dB |
20 x Log (X/REF) |
|
Minimum Execution time = |
3.9 ms |
|
RMS |
1-pole digital filter |
|
Computed rms of inputs |
|
Minimum Execution Time = |
2.7 ms |
|
STAT |
MEAN, SDEV computer for sample population (N-1) |
|
Minimum Execution Time = |
900 µs |
|
CTHRM2K (FTHRM2K) |
ºC (ºF) temperature conversion for 2.2 kW thermistor |
(HP 40653A) |
Minimum Execution Time = |
160 µs |
|
CRTD85 (FRTD85) |
ºC (ºF) temperature conversion for RTD of 100W, |
|
Alpha = |
0.00385 |
Minimum Execution Time = |
160 µs |
|
SCALE |
(X-OFFSET)/SCALE |
|
Minimum Execution Time = |
500 µs |
|
PFAIL |
Based on MIN, MAX registers |
3.9 ms |
Minimum Execution Time = |
|
|
dBm |
10 x Log [(X2/RES)/1 mW] |
|
Minimum Execution Time = |
3.9 ms |
|
FILTER |
1-pole digital filter |
|
Weighted Average of inputs |
|
Minimum Execution Time = |
750 µs |
|
CTHRM (FTHRM) |
ºC (ºF) temperature conversion for 5 kW thermistor (HP40653B) |
|
Minimum Execution Time = |
160 µs |
|
CTHRM10K (FTHRM10K) |
ºC (ºF) temperature conversion for 10 KW thermistor |
(HP406536) |
Minimum Execution Time = |
160 µs |
|
CRTD92 (FRTD92) |
ºC (ºF) temperature conversion for RTD of 100W, |
|
Alpha = |
0.003916 |
Minimum Execution Time = |
160 µs |
|
Front Panel Functions |
Null |
|
Stat |
|
Math Off |
|
Turns off all math functions |
|
Mean |
|
Displays the average of a number of readings within Statistics |
|
S Dev |
|
Displays the standard deviation for a range of readings within Statistics |
|
Upper/Lower |
|
Displays the highest or lowest of a range of readings within Statistics |
|
Scale |
|
Scales the reading up to 1 |
|
Synch ACV |
|
Turns on the most accurate acv mode |
|
Filter On/Off (LF OFF) |
|
Set to Off for synch ac > 50 kHz |
|
General Specifications |
Operating Temperature |
|
Operating Humidity Range |
|
Physical Characteristics |
Dimensions: |
88.9 mm H x 425.5 mm W x 502.9 mm D (3.5 in. H x 16.75 in. W x 19.8 in. D) |
Net Weight: |
12 kg (26.5 lbs) |
Shipping Weight: |
14.8 kg (32.5 lbs) |
|
IEEE-488 Interface |
Complies with the following: |
IEEE-488.1 Interface Standard |
|
IEEE-728 Codes/Formats Standard |
|
HPML (HP Multimeter Language) |
|
Storage Temperature |
|
Warm up Time |
|
4 hours to published specifications |
|
Power Requirements |
|
100/120V, 220/240V ±10% |
|
48-66 Hz, 360-420 Hz automatically sensed |
Fused: |
1.5 @ 115V or 0.5A @ 230V |
|
Designed in Accordance with |
Safety: |
IEC 348, UL1244, CSA |
EMI: |
FTZ 1046, FCC part 15-J |
Classification: |
Classified under MIL-T-28800D as Type III, Class 5, Style E, and Color R |
|
Warranty |
|
Input Terminals |
|
Gold-plated Tellurium Copper |
|