Biomedicína

Fluke Bm. Dale601E - Elektrický tester

Fluke Bm. Dale601E - Elektrický tester

Overview
The 601 Pro SeriesXL is an automated electrical-safety analyzer that meets stringent international standards for electrical-safety testing of hospital and laboratory electromedical equipment. The 601 Pro SeriesXL is also available with a built-in printer. (See “Models.”)

More details

Dostupnosť: Na objednávku
 
Informácie

Standard Features

  • EN60601-1, EN601010-1, and AAMI test loads (user selectable)
  • Multiple enclosure-leakage points
  • Multiple patient-applied-part types
  • Power ON/OFF delay
  • DC only for patient- and auxiliary-leakage tests
  • User-programmable test sequences
  • Manual, auto, step, and computer-control modes
  • ASCII data transfer
  • Memory for up to 1000 device-information records

Optional Features

  • Onboard thermal printing
Specifications
Voltage
(single and dual lead) 
Range:0 to 300 V True RMS
Accuracy:DC - 100 Hz ± 1.5 % of reading ± 1 LSD
Insulation Resistance
Range:0.5 to 400.0 MΩ
Accuracy:± 5 % of reading ± 2 LSD
Current Consumption
Range:0 to 15 A ac True RMS
Accuracy:± 5 % of reading ± 2 LSD
Mains on Applied Part
Applied Voltage:≥ 110 % of mains voltage
Accuracy:± 2 % of reading ± 6 µA
Supply Voltage
 90 to 265 Vac, autoswitching
Protective Earth Resistance
Range:0.000 to 2.999 Ω
Accuracy:± 5 % of reading ± 4 mΩ (1 A, 10 A, and 25 A test currents) (Refer to Operator’s Manual for additional specs qualifying the effects on accuracy of variations in load inductance and phase angle.)
IEC601-1 and AAMI Leakage Currents
Range:0 to 8000 µA True RMS
Accuracy:(per IEC601-1 or AAMI filter)
 - DC - 1 kHz ± 1 % of reading ± 1 µA
 - 1 to 100 kHz ± 2 % of reading ± 1 µA
 - 100 kHz to 1 MHz ± 5 % of reading ± 1 µA
DC-Only Frequency Response:DC - 5 Hz (approx)
IEC1010-1 Leakage Current
Range:0 to 16000 µA True RMS (For some leakage tests, displayed range may be 0 to 9999 when using IEC1010 test load.)
Accuracy:(per IEC1010-1 figure A.1 filter)
 - DC - 1 kHz ± 2 % of reading ± 2 µA
 - 1 to 100 kHz ± 4 % of reading ± 2 µA
 - 100 kHz to 1 MHz ± 10 % of reading ± 10 µA
DC-Only Frequency Response:DC - 5 Hz (approx)
ECG Simulation and Performance Testing
ECG Complex:30, 60, 120, 180, 240 BPM
Performance 
Pulse:30, 60 BPM, 63 ms pulse width
600 to 700 µs rise and fall time 
Sine Waves:10, 40, 50, 60, 100 Hz
Square Wave:0.125, 2.000 Hz (50 % duty cycle)
Triangle Wave:2 Hz, 2 mV
Arrhythmia
 Atrial fibrillation
 Atrial flutter
 Atrial tachycardia
 Idioventricular
 PVC1
 R on T
 Run
 Ventricular fibrillation
 Ventricular tachycardia
Performance:± 2 % of reading for rate or ± 5 % of reading for amplitude, fixed at 1 mV peak on a Lead II ECG connection (except for triangle wave, which is 2 mV peak-to-peak)
Dimensions
 16.62 in L x 11.75 in W x 5.56 in H
 (42.2 cm L x 30 cm W x 14.1 cm H )
Weight
 17 lb
 (7.7 kg)

Príslušenstvo

    30 Ďalších produktov v rovnakej kategórii

    Porovnať 0