Fluke Bm. DataSim 6100 - Simulátor pacienta

Fluke Bm. DataSim 6100 - Simulátor pacienta

The DataSim 6100 is a patient simulator designed to meet your clinical training needs.

More details

Dostupnosť: Na objednávku
 
Informácie
  • Interface with Resusci-Anne™, Arrhythmia Anne™, and Chris Clean™
  • Synchronization of hemodynamic waveforms
  • ECG/respiration
  • Manual PAC and PVC insertions
  • Swan-Ganz procedure
  • Training capabilities
  • Video adapter interface

Optional Features

  • Expansion modules
DataSim 6100 Patient Simulator

Specifications

Specifications
Power
AC power provided by (choice of) 115 V or 230 V charger (standard with instrument)
Battery Type:12 V, 1.9 AH sealed lead acid rechargeable
Battery Capacity:ten hours
Battery Charge Time:5 to 95 % of complete charge in 14 hours
ECG
Output signals
- High Level:1 V/1 mV
Heart rate
- Range:30 to 300 BPM
- Accuracy:± 1 %
Output connectors
- Low Level:12-Lead electrode snaps
- High Level:Switchcraft 15GM7F
- Pacemaker Artifact:-8 mV, 1 ms
- Sample Rate:250 sample/s max
Performance testing
- Linearity:2.5 Hz triangular wave
- HR Cal Check:30 to 300 BPM
- Chart Speed:2.5 Hz square wave
- Amplitude:0.25, 0.50, 1.00, 1.50, 2.00, and 2.50 mV
Blood Pressure
Output signals
- High Level:1V/100 mmHg
- Transducer:5 and 40 uV/VmmHg
- Exciter Voltage:10 Vac or dc max
- Static Range:0 to 250 mmHg, adjustable in 5 % increments
- Accuracy:± 5 % or 1 mmHg, whichever is greater
- Waveforms:Static, square wave, and physiological dynamic
Output connector
- Switchcraft15GM7F
Performance testing
- Static:0 to 250 adjustable in 5 % increments
- Square Wave:0 to 250 mmHg adjustable in 5 % increments
Respiration
Output signals
- Delta Impedance:0.25 to 2.50 Ω Lead I
- Base Impedance:250 Ω Lead I, 750 Ω Lead II
- Rate Range:0 to 80 BPM
- Lead:All leads
Performance testing
- Rate:0, 10, 20, 40, and 80 per min
- Delta Impedance:0.25, 0.50, 1.00, 1.50, 2.00, and 2.50 Ω
- Coincidence check
General (six channels)- ECG/arrhythmia, respiration, arterial pressure, PA pressure, RA pressure, and auxiliary channel for optional cardiac output, CO2, and other parameters.
Fundamental Rhythms and Sequences
Normal sinus rhythm
Sinus tachycardia and bradycardia
Ventricular and sinus asystole
Atrial tachycardia
Atrial flutter and fibrillation
AV blocks (1st degree, 2nd degree Mobitz I and II, and 3rd degree)
Unifocal and multifocal PVCs
Ventricular tachycardia
Ventricular fibrillation
PVCs at 1 to 35/min for any rhythm
Couplet
Triplet
Bigeminy
Junctional
Accelerated junctional
PACs and PJCs
Atrial tachycardia with aberrant conduction
Idioventricular
Agonal
ST-segment elevation and depression
Pacemaker (atrial, ventricular, and AV sequential)
Failure to capture and sense
Bundle-branch block
Cardiac-failure sequence
Conversion sequence
Dimensions
13 in H x 10 in W x 4.7 in D
(33 cm H x 25.4 cm W x 11.94 cm D)
Weight
7.05 lb
(3.2 kg)