Biomedicína

Fluke Bm. ESA601 - Tester bezpečnosti zdravotníckych zariadení

Fluke Bm. ESA601 - Tester bezpečnosti zdravotníckych zariadení

ESA601 Electrical Safety Analyzer tests medical equipment to both US and international standards. Users simply flip a switch to change between AAMI or IEC electrical safety testing load. The US version includes overlays in AAMI or IEC nomenclature so technicians use the terms that are most familiar to them. Multiple language overlays, outlets, and power cords are available for convenient use in many countries.

Ten applied part connections allow for lead-to-ground (patient), lead-to-lead (patient auxiliary), and lead isolation (mains on applied part) leakage testing of equipment with multiple applied parts.

Designed for on-the-go testing, the portable analyzer is lightweight and compact and comes with a sturdy handle for easy carrying.

For an automated solution, Fluke Biomedical’s Ansur software plug-in for the ESA601 allows technicians to use a PC to run autosequences, document results, and print reports.

More details

Dostupnosť: Na objednávku
 
Informácie
  • Selectable AAMI or IEC test load
  • Ten applied parts-lead connectors
  • 90 V (min) to 264 V (max) autoswitching power supply
  • Dual-lead leakage and dual-lead voltage tests
  • Easy-to-read display
  • Compact and portable
  • Ansur plug-in software available to automate testing and document results
  • Multiple outlets and power cords for compatibility in multiple countries
  • Overlays in English, German, French, and Italian
  • Overlay for USA version with either IEC or AAMI nomenclature
  • RS-232 serial port for PC control and printing
Biomedical Test   >   Electrical Safety Analyzers   >  
ESA601 Electrical Safety Analyzer
Specifications
Power
Mains Power Operating Range:90 min to 264 max V ac (47 Hz to 63 Hz), autoswitching
Power Ratings:16 A at 264 V max
Mains Inlet
The instrument uses a standard IEC 60320-1/C20 mains inlet rated at 16 A and 250 V for Class-1 equipment in cold conditions.
Equipment Outlet
(See “Models.”)
Voltage Measurements
Autoranging:Yes
Display Resolution:1 V
Range (Mains Voltage):90 V to 264 V True RMS
Range (Accessible Voltage):0 V to 264 V True RMS
Accuracy:± 2 % of reading ± 2 V
Frequency Response:DC to 1000 Hz (-3 dB point)
Crest Factor:< 3
DC Input Impedance:1 MΩ
Earth-Resistance Measurement
Autoranging:Yes
Display Resolution:0.001 Ω
Range:0.000 Ω to 1.999 Ω
Accuracy:± 2 % of reading ± 5 MΩ
Current Source Amplitude:1 A dc (± 10 %)
Insulation Measurement
Auto-ranging:Yes
Display resolution:0.1 MΩ
Range:0.5 M to 100 MΩ
Accuracy:0.5 MΩ to 20 MΩ, ± 2 % of reading, ± 200 kΩ (above 20 MΩ, ± 5 % of reading, ± 200 kΩ Voltage source amplitude: 500 V dc (± 10 %)
VDE Differential Current
Auto-ranging:Yes
Display resolution:10 µA
Range:10 µA to 10000 µA
Accuracy:± 2 % of full scale
Leakage-Current Measurements
Autoranging:Yes
Display Resolution:1 μA
Configuration:RMS current flowing through the ANSI/AAMI ES1 1993 or IEC60601-1 test load
Measurement Range:0 to 8000 μA True RMS
Accuracy:± 1 % of reading ( ± 2 μA) @ dc and 25 Hz to 1000 Hz*; ± 2.5 % of reading (± 2 μA) 1 kHz to 200 kHz
*Full-scale input: Accuracy of mains-on-applied-part leakage currents: ± 2 % of reading ± 6 μA
Measururing instrument frequency responseDC to 1 MHz (-3 dB)
Crest Factor:< 3
Input impedance:Per Figure 15 of IEC601-1 or AAMI ES 1
Applied-Part Connections
The instrument contains ten banana jacks that serve as the applied-part connectors.
RS-232 Serial Port
The ESA601 utilizes a null modem serial cable connection to its RS-232 port. This port is used(1) to provide test-result outputs (sent in ASCII format when the PRINT pushbutton is pressed) to a computer or serial ASCII printer;
(2) to download processor firmware; and
(3) to provide an interface for remote operation of the ESA601.
Protection Circuitry
- All signal I/O connections, excluding the RS-232 port and earth connections, withstand a continuous input of 264 Vac, 47 to 63 Hz, or ± 264 Vdc without causing permanent damage. (Blowing a fuse does not cause permanent damage.)
- The mains input is protected by a double-pole circuit breaker.
Dimensions
9.8 in L x 8.2 in W x 3.0 in H
(24.0 cm L x 21.1 cm W x 7.6 cm H)
Weight
5.2 lb
(2.4 kg)