Specifications
Specifications |
Power |
|
AC power provided by (choice of) 115 V or 230 V charger (standard with instrument) |
Battery Type: |
12 V, 1.9 AH sealed lead acid rechargeable |
Battery Capacity: |
ten hours |
Battery Charge Time: |
5 to 95 % of complete charge in 14 hours |
|
ECG |
Output signals |
|
- High Level: |
1 V/1 mV |
Heart rate |
|
- Range: |
30 to 300 BPM |
- Accuracy: |
± 1 % |
Output connectors |
|
- Low Level: |
12-Lead electrode snaps |
- High Level: |
Switchcraft 15GM7F |
- Pacemaker Artifact: |
-8 mV, 1 ms |
- Sample Rate: |
250 sample/s max |
Performance testing |
|
- Linearity: |
2.5 Hz triangular wave |
- HR Cal Check: |
30 to 300 BPM |
- Chart Speed: |
2.5 Hz square wave |
- Amplitude: |
0.25, 0.50, 1.00, 1.50, 2.00, and 2.50 mV |
|
Blood Pressure |
Output signals |
|
- High Level: |
1V/100 mmHg |
- Transducer: |
5 and 40 uV/VmmHg |
- Exciter Voltage: |
10 Vac or dc max |
- Static Range: |
0 to 250 mmHg, adjustable in 5 % increments |
- Accuracy: |
± 5 % or 1 mmHg, whichever is greater |
- Waveforms: |
Static, square wave, and physiological dynamic |
Output connector |
|
- Switchcraft |
15GM7F |
Performance testing |
|
- Static: |
0 to 250 adjustable in 5 % increments |
- Square Wave: |
0 to 250 mmHg adjustable in 5 % increments |
|
Respiration |
Output signals |
|
- Delta Impedance: |
0.25 to 2.50 Ω Lead I |
- Base Impedance: |
250 Ω Lead I, 750 Ω Lead II |
- Rate Range: |
0 to 80 BPM |
- Lead: |
All leads |
Performance testing |
|
- Rate: |
0, 10, 20, 40, and 80 per min |
- Delta Impedance: |
0.25, 0.50, 1.00, 1.50, 2.00, and 2.50 Ω |
- Coincidence check |
|
General (six channels) |
- ECG/arrhythmia, respiration, arterial pressure, PA pressure, RA pressure, and auxiliary channel for optional cardiac output, CO2, and other parameters. |
|
Fundamental Rhythms and Sequences |
|
Normal sinus rhythm |
|
Sinus tachycardia and bradycardia |
|
Ventricular and sinus asystole |
|
Atrial tachycardia |
|
Atrial flutter and fibrillation |
|
AV blocks (1st degree, 2nd degree Mobitz I and II, and 3rd degree) |
|
Unifocal and multifocal PVCs |
|
Ventricular tachycardia |
|
Ventricular fibrillation |
|
PVCs at 1 to 35/min for any rhythm |
|
Couplet |
|
Triplet |
|
Bigeminy |
|
Junctional |
|
Accelerated junctional |
|
PACs and PJCs |
|
Atrial tachycardia with aberrant conduction |
|
Idioventricular |
|
Agonal |
|
ST-segment elevation and depression |
|
Pacemaker (atrial, ventricular, and AV sequential) |
|
Failure to capture and sense |
|
Bundle-branch block |
|
Cardiac-failure sequence |
|
Conversion sequence |
|
Dimensions |
|
13 in H x 10 in W x 4.7 in D |
|
(33 cm H x 25.4 cm W x 11.94 cm D) |
|
Weight |
|
|