Fluke Bm. MPS450 - Simulátor pacienta "ďalšej generácie"

Fluke Bm. MPS450 - Simulátor pacienta "ďalšej generácie"

Tento prístroj sa už nedodáva. Pozrite si prosím jeho náhradu.

 

 

 

 

The MPS450 is Fluke Biomedical"s next-generation, portable, multiparameter patient simulator for your comprehensive testing and training needs. Cardiac Output and Fetal/Maternal ECG models are also available. (See “Models.”)

More details

Dostupnosť: Nedostupné
 
Informácie
  • ECG simulation/12-lead independent outputs
  • 43 arrhythmia selections
  • Four invasive BP channels including Swan-Ganz
  • Respiration simulation
  • Temperature simulation
  • BP synchronization with ECG
  • Large, bright 4-line by 20-character super twist display
  • R-Wave detection test
  • 9 V battery operation
  • External battery tests (9 V)
  • RS-232 port for computer control
  • Compact size
  • Front-panel soft-key user interface
  • Universal ECG connectors
  • High-level ECG output
  • Expansion port
Biomedical Test   >   Simulators & Controllers   >   Multiparameter Patient Simulators   >  
MPS450 Patient Simulator
Specifications
Power
 Two 9 V alkaline batteries (eight hours min continuous power).
 Optional battery eliminator
Normal-Sinus-Rhythm Waveform
ECG Reference:ECG amplitudes specified for Lead II (calibration), from the baseline to the peak of the R-Wave; other leads proportional
Normal Sinus Rhythm:12-lead configuration with independent outputs referenced to right leg (RL); output to ten Universal ECG Jacks, color-coded to AHA and IEC Standards
Low-Level Amplitude:0.05 to 0.50 mV (0.05-mV steps); 0.5 to 5.5 mV (0.5-mV steps) (Power-on default: 1 mV)
Amplitude Accuracy:± 2 % of setting Lead II
High-Level Output:(available on the BP3 connector) 0.2 V/mV ± 5 % of the ECG-amplitude setting
ECG Rate:30, 40, 45, 60, 80, 90, 100, 120, 140, 160, 180, 200, 220, 240, 260, 280, and 300 BPM (Power-on default: 80 BPM)
Rate Accuracy:± 1 % of setting
ECG-Waveform Selection:Adult (80 ms) or pediatric (40 ms) QRS duration (Power-on default: adult QR/P-R interval of 0.16 seconds)
Superimposed Artifact:50 and 60 Hz, muscle, baseline wander, respiration
ST-Segment Elevation/Depression:Adult mode only, -0.8 mV to +0.8 mV, in 0.1-mV steps; and additional steps, +0.05 mV and -0.05 mV (Power-on default: elevation 0 mV)
Pacemaker Waveform
Pacer-Pulse Amplitude:1, 2, 5, 10 mV, ± 10 % (Power-on default: 5 mV)
Pacer-Pulse Width:0.1, 0.5, 1.0, 1.5, 2.0 ms, ± 5 % (Power-on default: 1.0 ms)
Pacing Rate:75 BPM
Waveforms:(Power-on default: atrial waveform)
 - Atrial
 - Asynchronous 75 BPM
 - Demand with frequent sinus beats
 - Demand with occasional sinus beats
 - AV sequential
 - Noncapture (one time)
 - Nonfunction
Arrhythmias
Baseline NSR:80 BPM
PVC Focus:Left focus, standard timing (except where specified)
Supraventricular Arrhythmias:(Power-on default: atrial fibrillation, coarse)
 - Atrial fibrillation, coarse
 - Atrial fibrillation, fine
 - Atrial flutter
 - Sinus arrhythmia
 - Missed beat (one time)
 - Atrial tachycardia
 - Paroxysmal atrial tachycardia (PAT)
 - Nodal rhythm
 - Supraventricular tachycardia
Premature Arrhythmias (all one-time events):(Power-on default: premature atrial contraction)
 - Premature atrial contraction (PAC)
 - Premature nodal contraction (PNC)
 - PVC1 left ventricular
 - PVC1 left ventricular, R on T
 - PVC2 right ventricular
 - PVC2 right ventricular, early
 - PVC2 right ventricular, R on T
 - Multifocal PVCs
Ventricular Arrhythmias:(Power-on default: PVCs 6/min)
 - PVCs 6/min (Power-on default)
 - PVCs 12/min
 - PVCs 24/min
 - Frequent multifocal PVCs
 - Bigeminy
 - Trigeminy
 - Multiple PVCs (one-time run of 2PVCs)
 - Multiple PVCs (one-time run of 5 PVCs)
 - Multiple PVCs (one-time run of 11 PVCs)
 - Ventricular tachycardia
 - Ventricular fibrillation, coarse
 - Ventricular fibrillation, fine
 - Asystole
Conduction Defects:(Power-on default: first-degree heart block)
 - First-degree heart block
 - Second-degree heart block
 - Third-degree heart block
 - Right-bundle-branch block
 - Left-bundle-branch block
ECG-Performance Testing
Amplitude:0.05 to 0.50 mV (0.05-mV steps); 0.5 to 5.5 mV (0.5-mV steps) (Power-on default: 1 mV)
Pulse Wave:30, 60 BPM, with 60 ms pulse width
Square Wave:2, 0.125 Hz (Power-on default: 2 Hz)
Triangle Wave:2, 2.5 Hz
Sine Wave:0.5, 5, 10, 40, 50, 60, 100 Hz
R-Wave-Detection Waveform:Haver-Triangle
R-Wave Rate:30, 60, 80, 120, 200, and 250 BPM (Power-on default: 60 BPM)
R-Wave Width:20 to 200 ms (10-ms steps); additional steps, 8, 10, and 12 ms (Power-on default: 10 ms)
Rate Accuracy:± 1 %
Amplitude Accuracy:± 2 %, Lead II (exception, ± 5 % for R-Waves ≤ 20 ms)
Respiration
Rate:0 (OFF), 15, 20, 30, 40, 60, 80, 100, 120 BrPM (Power-on default: 20 BrPM)
Impedance Variations (∆Ω):0.2, 0.5, 1, or 3 Ω (Power-on default: ∆ 1 Ω )
Accuracy Delta:± 10 %
Baseline:500, 1000, 1500, 2000 Ω, Leads I, II, III (Power-on default: 1000 Ω)
Accuracy Baseline:± 5 %
Respiration Lead:LA or LL (Power-on default: LA)
Apnea Selection:12, 22, or 32 s (one-time events), or continuous (Apnea ON = respiration OFF) (Power-on default: 12 s apnea)
Blood Pressure
Input/Output Impedance:300 Ω ± 10 %
Exciter-Input Range:2 to 16 V RMS
Exciter-Input-Frequency Range:DC to 5000 Hz
Transducer Sensitivity:5 or 40 µV/V/mmHg (Power-on default: 5 µV/V/mmHg)
Pressure Accuracy:± (2 % of setting + 2 mmHg)
Static Levels, P1: -10, 0, 80, 160, 240, 320, 400 mmHg (Power-on default: 0 mmHg)
Static Levels, P2: -10, 0, 50, 100, 150, 200, 240 mmHg (Power-on default: 0 mmHg)
Static Levels, P3: -5, 0, 20, 40, 60, 80, 100 mmHg (Power-on default: 0 mmHg)
Static Levels, P4: -5, 0, 20, 40, 60, 80, 100 mmHg (Power-on default: 0 mmHg)
Dynamic Waveforms, P1: 
- Arterial:120/80
- Radial artery:120/80
- Left ventricle:120/00
- Right ventricle:25/00
Dynamic Waveforms, P2: 
- Arterial:120/80
- Radial artery:120/80
- Left ventricle:120/00
- Right atrium (central venous or CVP):15/10
- Right ventricle:25/00
- Pulmonary artery:25/10
- Pulmonary-artery wedge:10/2
- Left atrium:14/4
Dynamic Waveforms, P3: 
- Arterial:120/80
- Radial artery:120/80
- Left ventricle:120/00
- Right atrium (central venous or CVP):15/10
- Right ventricle:25/00
- Pulmonary artery:25/10
- Pulmonary-artery wedge:10/2
- Left atrium:14/4
Dynamic Waveforms, P4 (Swan-Ganz Sequence):Right atrium (CVP); right ventricle RV); pulmonary artery (PA); pulmonary-artery wedge (PAW)
Respiration Artifact:BP delta change from 3 to 16 mmHg
BP Output:Mini DIN 7-Pin
Temperature
Temperature:0 °C (32 °F), 24 °C (75.2 °F), 37 °C (98.6 °F), and 40 °C (104 °F) (Power-on default: 0 °C / 32 °F)
Accuracy:± 0.1 °C
Compatibility:Yellow Springs, Inc. (YSI) Series 400 and 700
Output:mini DIN 7-pin
Cardiac Output (optional)
Catheter Type:Baxter Edwards, 93a-131-7f
Calibration Coefficient:0.542 (0 °C injectate), 0.595 (24 °C injectate)
Blood Temperature:37 °C (98.6 °F) ± 2 %
Injectate Volume:10 cc
Injectate Temperature:0 °C or 24 °C ± 2 % value (Power-on default: 0 °C injectate)
Cardiac Output:2.5, 5, 10 lpm ± 5 % (Power-on default: 2.5 lpm)
Faulty-Injection Curve:(Waveform for simulation available)
Left-to-Right-Shunt Curve:(Waveform for simulation available)
Calibrated Pulse:1.5 ° for 1 second (37 ° * 35.5 °) (Waveform for simulation available)
Repeatability:± 1 %
Fetal/Maternal ECG (optional)
Maternal Heart Rate (Fixed):80 BPM
Fetal Heart Rate (Selectable):60, 90, 120, 140, 150, 210, and 240 BPM (Power-on default: 120 BPM)
Fetal Heart Rate (IUP):140 BPM at beginning, then varying with pressure
Intrauterine-Pressure Waveforms:- Uniform acceleration (140-175-140 BPM; rate change lagging IUP contraction by 30 s)
 - Uniform deceleration (140-100-140 BPM; rate change lagging IUP contraction by 30 s) (Power-on default)
 - Early deceleration (140-100-140 BPM; no IUP lag time)
 - Late deceleration (140-100-140 BPM, starting at IUP peak)
Wave Duration:90 seconds, bell-shaped pressure curve, from 0 to 90 mmHg and returning to 0, ± 4 mmHg (max)
IUP Period:2, 3, or 5 minutes; and manual (Power-on default: manual)
Computer Setup
Port:Bidirectional (Data Communications Equipment) RS232
Baud Rate:9600
Parity:None
Stop Bits:1
Data Bits:8
Battery Replacement
 Warning for low-battery condition (batteries to be replaced at this time)
Temperature
Storage:-25 to 50 ºC
Operation:10 to 40 ºC
Humidity
 80 % max relative humidity
Dimensions
 7.5” L x 7.25” W x 2” H
 (19 cm L x 18.415 cm W x 5 cm H)
Weight
 1.35 lb
 (.612 kg)