~Nedodávané a ukončená výroba

TandD RTR-51L/51AL

TandD RTR-51L/51AL

RTR-5 Series - Wireless Compact Waterproof Data Logger

Our Revolutionary Wireless RTR-5 Series is a system wherein the data measured and recorded by the data logger units can be transmitted to any of our handy multi-functional Communication Interfaces (RTR-57U / RTR-50 / RTR-5W) via short wave radio communication. (For more details about how to use and functions included, please click on the figures below.)
The transmission range is up to 100 meters if direct and unobstructed.
The RTR-5 Series gives you the power to check data, start and stop recording and collect data on-site without ever having to gather the units.

More details

Dostupnosť: Nedostupné
 
Informácie

Model Names "RTR-51A, RTR-52A and RTR-53A" are the current names for CE approved models and RTR-51, RTR-52, and RTR-53 are the current names for FCC approved models.
Model Names "RTR-52Pt" and "RVR-52A" are used for both CE and FCC approved models.

NOTE:
Although former CE models with names “RTR-51, RTR-52, RTR-53 and RVR-52” do not include an adjustment function, all other specifications are the same.

Wireless Data Communication Function
Our RTR-51/51A/52/52A/52Pt/53/53A and RVR-52A Data Logger units utilize the RTR-57U to collect recorded temperature data from the units via our exclusive short-wave wireless technology. The RTR-57U also allows you to wirelessly control recording settings and start recording, as well as other functions of the Data Loggers. Moreover, the Data Loggers can wirelessly transmit measurements in the wide range of -30 to 80°C from distances of up to 100 meters if unobstructed and direct.



Note: If collecting data via wireless communication, it is necessary to register via computer the RTR-51/51A/ 52/52A/52Pt/53/53A and RVR-52A as Remote Units and the RTR-57U as the Base Unit. The data recorded by the Remote Units can be downloaded via optical communication by placing it on the communication port of the RTR-57U.

Register up to 3840 Units on 1 RTR-57U
RTR-57U is desighed to manage any combination of RTR-51/51A/52/52A/52Pt/53/53A and RVR-52A units in groups.
One RTR-57U unit can be set up to manage up to 60 groups, with each group containing up to 64 units. If being set up via computer each RTR-57U can be set to handle 15 groups with each group containing 250 units.

Compact, Durable, Water Resistant
The lightweight yet durable water resistant construction allows you to use this unit under the harshest of conditions. It can be reliably used in high condensation areas, refrigerated and frozen environments as well as underground.

Note: The waterproofing is sufficient for most situations, but is not for continued immersion.

Wireless Communication Possible in Range of -30 to 80°C
Because wireless communication for all RTR-51/51A/52/52A/52Pt/53/53A and RVR-52A units can be made within the wide range of -30 to 80°C, it is possible to use the units to manage temperature in below freezing conditions such as frozen transportation and storage.

Recording Capacity: 16,000 Readings
One RTR-51/51A or RTR-52/52A unit can record up to 16,000 readings. If set at a one hour recording interval that gives you 666 days or almost two years of recording.

Adjustment Function
Can be used only with the RTR-51/51A/52/52A/52Pt/53/53A and RVR-52A. This function allows you to correct or adjust for errors in measurement values. The Data Logger will record only the adjusted value.

Note: We cannot guarantee that after carrying out adjustment the measuring accuracy will improve for all measuring ranges.
The former CE models with names "RTR-51, RTR-52, RTR-53" and RVR-52" do not include an adjustment function.

Monitoring Function
Our RTR-5 Series allows you to wirelessly monitor current readings of one group of RTR-5 Series units, with the readings being sequentially displayed on the handheld RTR-57U Unit.

Monitoring intervals can be set on the RTR-57U for 15 and 30 seconds as well as from 1 minute to 60 minutes for up to 64 units in one group. The high-speed communication allows you to collect 64 units worth of data in just 20 seconds. Moreover, the downloaded data is sequentially displayed on the handheld unit for 2 seconds. At the same time, if a set upper or lower limit has been exceeded a warning will be displayed and an alarm sounded.



Note: If monitoring is carried out on a regular basis the battery life of the RTR-51/51A/52/52A/52Pt/53/53A units will be shortened. For example, if the monitoring interval is set at 1 minute the battery life will be about 4 months.

About 6 Months of Continuous Use with Lithium Battery
Using our specially designed low energy consumption circuit this unit can run on one lithium battery for up to six months of continued use. In addition, by using our specially designed optional large capacity battery pack, the unit can used continuously for up to two and a half years. No need to worry about where you place it, as the battery will allow you to measure and record over long periods of time whether the unit is in transit or in a distant place.

Note: Battery life will depend on the recording environment, recording interval, communication frequency, and ambient temperature. The above battery life test was carried out using brand new batteries and in no way do we guarantee a battery"s life.
Removing the battery and not replacing it will result in a loss of all recorded data.

Remote Control Command
Remote Control Command is possible to set up your computer to enable the control of downloading via wireless communication between an RTR-57U unit connected to your computer and any RTR-51/51A/52/52A/52Pt/53/53A and RVR-52A unit(s) within communication range. To use this function you will need to input a special command protocol into your computer"s software system. For more details about the set up and use of this function please contact the T&D sales office.

Wide Selection of Recording Intervals / Two Recording Methods
The RTR-51/51A/52/52A/52Pt/53/53A and RVR-52A give you 15 recording intervals (1 second to 1 hour) to choose from. Each unit allows you two choices of recording method:
One time method: when 16,000 readings have been reached, recording stops.
Endless method: when 16,000 readings have been reached, it automatically overwrites oldest data.

Built-in Pre-heat Function
The RVR-52A includes an internal pre-heat function which sends signals to turn ON / OFF external sensors, etc...in time with the starting of recording.

CE! models
( for EU )
RTR-51A/51ALRTR-52A/52ALRTR-52Pt/52PtLRTR-53A/53AL
FCC models
( for US )
RTR-51/51LRTR-52/52LRTR-52Pt/52PtLRTR-53/53L
Measurement Channel1Temperature Channel1Temperature Channel1Temperature Channel1Temperature and
1 Humidity Channel
Internal Temp. Sensor-40 to 80°C---
External Temp. Sensor--60 to 155°C-199.9 to 600°C(*1)0 to 55°C
Thermal-Constant Time15 minutes
L Type: 25 min.
Approx. 15 Sec. (in air)
Approx. 2 Sec. (in agitated water)
--
Sensor Response Time---About 7 min.
(90% response)
External Humidity Sensor---10 to 95%RH
Measurement Display Resolution0.1°C0.1°C0.1°C0.1°C/1%RH
Measurement AccuracyAvg±0.5°CAvg±0.3°C(-20 to 80°C) Avg±0.5°C(-40 to -20°C/80 to 110°C) Avg±1.0°C(-60 to -40°C/110 to 155°C)±0.3°C(-199.9 to 80°C) ±0.5°C(80 to 450°C) ±1.0°C(450 to 600°C) (at 0 to 50°C/Sensor accuracy is not included.)Avg±0.3°C ±5%RH (at 25°C and 50%RH)
Recording Interval1,2,5,10,15,20,30 seconds / 1,2,5,10,15,20,30,60 minutes Total of 15 choices (Excluding Event data)
Recording Capacity16,000 Readings x 1 Channel8,000 Readings x 2 Channnel
Recording ModeEndless (Overwrite oldest data when capacity is full) / One-time (Stop recording when capacity is full)
LCD Display ItemsCurrent Readings / Recording Settings / Battery Life Warning / Over Measurement Range Warning / Amount of Recorded Data
PowerLithium Battery (LS14250(SAFT)) x 1 / L Series (Large Capacity Battery Pack) :Litium Battery (LS26500) x 1 / AC Adaptor (sold separately) (*2)
Battery LifeApprox. 6 months / L Series: Approx 2 years and 6 months
Wireless SpecificationsETSI EN 300-220 [ Frequency Range: 433.050 to 434.79MHz ],
FCC Part15 Section249 / IC RSS-210 [ Frequency Range: 902 to 928MHz ]
Transmission RangeApprox. 100m (May vary with conditions)
DimensionsH62mm x W 47mm x D19mm(D46.5mm:L Series)
(excluding protrusions / antenna length 20mm)
InterfaceWireless Communication / Optical Communication
Communication SpeedWhen downloading data (Wireless) about 2,000 readings per minute [Collection of a full unit of data= about 420 seconds (optical communication= about 160 seconds)]
Water Resistance (*4)IP67 (immersion proof)IP64 (rated for use in daily life)
Standard Sensor / Input Cable-TR-5106 (length: Approx.0.6m)x 1RTR-05P1 (length: Approx.1m)x 1TR-3310 (length: Approx.1m)x 1
Unit Temp. Resistance-30 to 80°C ( Unit temp resistance and measurement range is -40 to 80°C but wireless communication cannot occur in an environment of less than -30°C )
OthersIn order to download data via wireless communication, it is necessary to purchase either an RTR-57U, RTR-5W, and/or RTR-50 unit. -For optical communication, TR-50C can also be used for all types of units.


RVR-52A/52AL
Measurement Channel1Voltage Channel1Pulse Channel1Event Channel
Input Voltage Range0 to 6.5V0 to 30V0 to 30V
Measurement Display Resolution1mV-1sec.
Measurement Accuracy±0.5%+5dgt
(at 0 to 40)
--
Detection Voltage-Lo: below 0.6V / Hi: above 2V
Response Time30Hz / Approx.15msec or more (over 2.5V)1 sec.
Response Polarity-Pu: select Lo-Hi or Hi-LoEv: Both Lo -Hi and Hi-Lo
Highest No. of Counts-Pu: 32,000Counts-
Input ImpedanceApprox. 1MPu, Ev: Pull-up Resistor Approx.100k
Recording Capacity16,000 Readings x 1 Channel16,000 Readings x 1 Channel8,000 Readings x 1 Channnel
Recording Interval1,2,5,10,15,20,30 seconds / 1,2,5,10,15,20,30,60 minutes
Total of 15 choices (Excluding Event data)
Recording ModeEndless (Overwrite oldest data when capacity is full) /
One-time (Stop recording when capacity is full)
LCD Display ItemsCurrent Readings / Recording Settings / Battery Life Warning /
Over Measurement Range Warning / Amount of Recorded Data
PowerLithium Battery (LS14250(SAFT)) x 1 /
L Series (Large Capacity Battery Pack) :Litium Battery (LS26500) x 1 /
AC Adaptor (sold separately)(*2)
Battery Life(*3)Approx. 6 months / L Series: Approx 2 years and 6 months
Wireless SpecificationsETSI EN 300-220 [ Frequency Range: 433.050 to 434.79MHz ],
FCC Part15 Section249 / IC RSS-210 [ Frequency Range: 902 to 928MHz ]
Transmission RangeApprox. 100m (May vary with conditions)
DimensionsH62mm x W 47mm x D19mm(D46.5mm:L Series)
(excluding protrusions / antenna length 20mm)
InterfaceWireless Communication / Optical Communication
Communication SpeedWhen downloading data (Wireless) about 2,000 readings per minute
[Collection of a full unit of data= about 420 seconds (optical communication= about 160 seconds)]
Water Resistance(*4)IP64 (rated for use in daily life)
Standard Sensor / Input CableRPR-7101 (length: Approx. 1.5m) x 1
Unit Temp. Resistance-30 to 80°C ( Unit temp resistance and measurement range is -40 to 80°C but wireless communication cannot occur in an environment of less than -30°C )
OthersThe RVR-52A includes an internal pre-heat function which sends signals to turn ON / OFF external sensors, etc...in time with the starting of recording.
Possible to measure current with the RVR-5203 4-20mA Probe.
In order to download data via wireless communication, it is necessary to purchase either an RTR-57U, RTR-5W, and/or RTR-50 unit.
-For optical communication, TR-50C can also be used for all types of units.
(*1)When measuring and recording using Fahrenheit, although it is possible to measure between -327.8 and 1112 degrees, the LCD on the Logger can only show between -199.9 and 999.9 degrees.
(*2)Contact your local authorized TandD sales representative.
(*3)Battery life depends on measurement environment, recording interval and battery performance.
(*4)Water protection is rated as splashproof (IP64) when the sensor is connected. Not for continued immersion.




1.Antenna
2.LCD Display
3.Optical Port
4.Back Case
5.Internal temperature sensor
6.Large Capacity Battery

Príslušenstvo

    30 Ďalších produktov v rovnakej kategórii

    Porovnať 0