Metrológia teploty

Hart Scientific 7900 - Regulátor (–100 do 400 °C )

Hart Scientific 7900 - Regulátor (–100 do 400 °C )

  • Všetky vlastnosti ako regulátor Hart 2100
  • Jednoduchá inštalácia

More details

Dostupnosť: Na objednávku
 
Informácie

Hart’s bath controllers have long been recognized as being among the finest in the world. They’re the most popular retrofit controller in the industry, and now they’re available for Rosemount baths. The Model 7900 Controller installs easily and can replace the Rosemount Model 915 for all Rosemount bath models.

This controller uses the same circuitry as Hart’s 2100 Controller to achieve long-lasting stabilities of ± 0.003 °C or better. Special noise-rejection techniques allow the 7900 to measure the very tiny resistance changes required for this level of stability. AC bridges are used within the controller to cancel thermal EMFs. Custom high-precision resistors contribute to short- and long-term stability and advanced filtering techniques force out troublesome line noise.

The Model 7900 includes a special circuit that monitors the controller’s microprocessor and automatically resets it if its operations are interrupted. Two separate cutout systems are also included for keeping your bath’s temperature within its normal range.

A software cutout uses an adjustable high-temperature limit that can be easily accessed through the front panel and set to match the requirements of your bath fluid. Should the control sensor measure a temperature beyond this upper limit, heating is shut down. If the bath’s temperature falls below its normal operating range, the heaters are turned on and the LN2 cooling is shut off. A second, independent hardware cutout monitors the bath’s temperature with a thermocouple and shuts down all heating and LN2 cooling if the bath’s temperature rises above its range.

These cutout features, combined with the superior reliability and long-term stability performance of the 7900, allow you to run your system for as long as you like between shut-downs—365 days a year, if you wish. Your bath can be ready for you to take measurements the minute you walk into your lab each day.

Specifications 

Temperature Control Range 

–100 °C to 400 °C 

Optional Ranges 

None 

Stability 

± 0.003 (± 0.001 typical) 

Stabilization Time 

30 minutes 

Display Accuracy 

± 1 °C 

Cooling Control 

LN2 – automatic 

Heating Control 

2-position, firmware or user controlled 

Firmware High-Temp Cutout 

Yes, volatile, programmable (independent of the controller) 

Hardware High-Temp Cutout 

Thermocouple controlled 

Memory 

Non-volatile; 8 programmable set-points, each with ramp and soak features 

Programmable Soak Time 

1 to 500 minutes 

Control Sensor 

100-ohm PRT;
alpha = 0.00385 

Interface 

RS-232 and IEEE standard 

Software 

Interface-it 

Operating Temperature 

5 °C to 50 °C 

Operating Voltage 

115 VAC (±10 %), 60 Hz 

CE Mark 

Contact Hart 

Current Rating 

20 amps max. 

Dimensions (WxHxD) 

311 x 114 x 279 mm
(12.25 x 4.5 x 11 in) 

Weight 

4 kg (9 lb.) 

Installation 

Freestanding or rack mounted with optional hardware 

Príslušenstvo

    30 Ďalších produktov v rovnakej kategórii

    Porovnať 0