| Specifications |
| Power |
| |
120/250 Vac, 50 W average, 100 W peak, 50/60 Hz |
|
| Analog Outputs |
| Cuff Pressure: |
0 to 499.95 mmHg FS ± 1.0 % of reading, 10 mV/mmHg |
| Pulse Pressure: |
0.0 to 5.0 mmHg FS ± 1.0 of reading, 1.0 V/mmHg |
|
| Digital Manometer |
| Pressure Range: |
- 499.75 to 499.75 mmHg |
| Measurement Parameters: |
instantaneous and peak |
|
| Displayed Graphics |
| |
Dynamic real-time NIBP cuff-pressure waveform programmed peripheral pulse and envelope waveforms |
|
| Display |
| |
Alphanumeric Graphic Display (LCD) |
| Alphanumeric Mode: |
8 lines x 40 characters |
| Graphics Mode: |
64 vertical x 240 horizontal dot matrix, backlight with viewing angle adjustment |
|
| Digital Interfaces |
| RS232C/Serial: |
bidirectional. Downloads cuff measurement data and controls test features with a compatible computer or via the medTester 5000C with the medCheck option. |
| Parallel Printer: |
Centronics compatible |
| Pulse Sync: |
0 V dc to 5 V dc (TTL) |
|
| Cuff Mandrel |
| Interlocking Plastic Blocks: |
four cuff circumferences, including 39.5 cm (large adult), 33 cm (adult), 26.6 cm (small adult) and 20 cm (child); maximum cuff : 15.25 cm. |
| Truncated Plastic Cylinders: |
Three neonatal cuff circumferences, including 14 cm, 10 cm, and 7.6 cm; maximum cuff width of 7.6 cm. |
|
| Pop-off Valve Testing |
| |
Automatic test for operation of the monitor’s relief valve |
| Measurement Parameters: |
instantaneous and peak pressure |
| Maximum Pressure: |
499.75 mmHg |
|
| System Leak Testing |
| Start Pressure: |
499.75 mmHg max |
| Elapsed Time: |
60 s (fixed) |
| Leak-Rate Range: |
0.25 mmHg/min to 499.75 mmHg/min |
| Pump: |
2.0 liters/minute minimum (free flow) |
|
| Accuracy |
| Systolic/Diastolic Mean Arterial Pressure (MAP): |
1 % of target value |
| Cuff Pressure: |
± 1 % of reading ± 1 mmHg |
| Input Overpressure Limit: |
± 1500 mmHg |
|
| Autosequences |
| |
Up to five user-programmable sequences to test NIBP monitors with a specific series of CuffLink performance tests, including static pressure test, leak test, and pop-off test |
| |
Up to eight adult-neonatal-arrhythmia dynamic NIBP selections, each of which can be cycled up to 99 times during the sequence. |
| |
Printable test report. |
|
| Displayed Real-Time Parameters |
| Instantaneous Cuff Pressure: |
0 mmHg to 300 mmHg |
| Peak Cuff Pressure: |
0 mmHg to 500 mmHg peak |
| Inflate/Deflate Time: |
0.1 s to 999.9 s |
| Inflate/Deflate Rate: |
0.1 mmHg/s to 999.9 mmHg/s |
| Total Measurement Time: |
0 s to 999.9 s max |
| Selected Heart Rate: |
selected systolic/diastolic (mean) target values |
| |
User-programmed vertical and horizontal shifts |
|
| Dynamic Noninvasive Blood Pressure |
| |
Simulation of a range of normal, hypertensive and hypotensive dynamic noninvasive blood pressures for typical adult, infant, and neonatal patients. Generation of normal, bradycardia, and tachycardia rhythm selections with a wide range of user-programmable peripheral pulse amplitudes (weak, normal and strong). Compatible with oscillometric NIBP devices. |
| Preprogrammed Target Value Selections: |
Adult Systolic/Diastolic (MAP) (mmHg): 60/30 (40), 80/50 (62), 400/65 (75), 120/80 (90), 150/100 (115), 200/150 (165) and 255/195 (215) |
| Neonatal/Pediatric Systolic/Diastolic: |
- Above selections, excluding 255/195 and 200/150 |
| Repeatability: |
± 1 % of selected target value |
| Adult Arrhythmia Selections: |
Baseline NIBP Target Value: (120/80) (NSR) |
| |
Atrial fibrillation (A-Fib) |
| |
Premature atrial contraction (PAC) |
| |
Premature ventricular contraction (PVC) |
| |
Missed beat (MB) |
| |
Aberrant sinus conduction (AS) |
| Preprogrammed Peripheral Pulse Waveforms: |
Pulse amplitude at MAP: 2.0 mmHg (typical adult value) |
| |
Pulse volume range: 0.0 to 5.1 ml |
| |
Pulse rise time: 80 ms (min) |
| |
Heart rates (adult & neonate): 30, 40, 60, 80, 120, 160, 200 & 240 BPM |
| |
Heart-rate accuracy: ± 1.0 % of selected rate |
| User-Programmable Target Value Shifts: |
Horizontal Axis: preprogrammed target value selections shifted in 1.0 mmHg steps over a maximum range of ± 100 mmHg to increase or decrease dynamic pressure values |
| |
Vertical Axis: relative amplitude shifted in 1 % increments over a maximum range from 0 to 200 % to simulate weak, normal, and strong peripheral pulses |
|
| Dimensions (WxDxH): |
| |
31.75 cm x 38.1 cm x 12.7 cm |
| |
(12.5 in x 15 in x 5 in) |
|
| Weight |
|