Fluke Bm. RF303 - Tester elektrochirurgických zariadení

Značka: Fluke

Electrosurgery-device testing has never been easier than with the RF303RS. User-selectable loads (50 to 750 Ω), signal averaging, battery power and portability make the RF303RS a very versatile tool for determining performance of multiple ESU “makes” and models.

Dostupnosť Na objednávku
Môžeme doručiť do:
16.8.2024
Možnosti doručenia
Kód: 1329
€1 €0,83 bez DPH
Fluke Bm. RF303 - Tester elektrochirurgických zariadení
Značka: Fluke
Fluke Bm. RF303 - Tester elektrochirurgických zariadení
Ukončená výroba
Doprava zadarmo pre každý nákup nad 80€ s DPH vytvorený cez náš eshop s doručením na Slovensko.
Doprava zadarmo
pre každý nákup nad 80€ s DPH vytvorený cez náš eshop s doručením na Slovensko.
Široké možnosti platby na dobierku, kartou online alebo bankovým prevodom.
Široké možnosti
platby a doručenia pre vaše pohodlie. Prijímame platobné karty, Google Pay alebo Apple Pay.
Tovar skladom odosielame v deň objednávky (do 13:00)
Tovar skladom
odosielame v deň objednávky kuriérom SPS alebo službou Packeta (objednávky do 13:00).
Záručný, pozáručný servis teplomerov, multimetrov a iného metrologického vybavenia.
Servis a kalibrácia
teplomerov, multimetrov a iného metrologického vybavenia vrátane záručného a pozáručného servisu.
  • New Signal Averaging Mode (SAMTM)
  • New RS232 port
  • Battery powered for completely isolated measurement
  • Digital display
  • HF-leakage-current measurement
  • 4-digit numeric LCD with backlight and power-save mode

Optional Features

  • All existing units upgradeable to the NEW RF303RS model
RF303 RS Electrosurgery Analyzer
Specifications
Modes of Operation
Line powered, battery powered, offline (battery maintenance charge)
Test Parameters
Power (W), HF/RF current (mA), test load (Ω)
Test-Load Selection
Monopolar and bipolar
Number of Selections: 15
Range: 50 Ω to 750 Ω
Step Size: 50 Ω
Accuracy: ± 4 % of selected value at 500 kHz
Current Measurement (Leakage)
Range: 30 mA to 2500 mA RMS
Resolution: 1 mA
Accuracy: ± 2.5 % of reading or ± 15 mA, whichever is greater
Power Measurement (Output)
Range: 1 to 400 W
Resolution: 0.1 W
Accuracy: ± 5 % of reading or ± 3 W, whichever is greater
Frequency Response
System Response: -3 dB points, 1 kHz to 10 MHz at 300 Ω
CQM Test
50 Ω to 750 Ω, 50-Ω steps (See “Test-Load Selection” for additional specifications)
Oscilloscope Output
Transformer coupled output, uncalibrated
Connector Type: BNC
Battery
Type: Sealed lead-acid
Time Between Recharge: Two hours (continuous use)
Time to Full Charge: Eight hours
Number of Cycles: 200
Top Panel Input Connectors
Connector Type: 4 mm-diameter safety sockets
Input Voltage Limit: 10,000 V peak
Input CurrentLlimit: 3 A RMS
Dimensions
6 in H x 13.25 in W x 11.5 in D
(15.2 cm x 33.7 cm x 29.2 cm)
Weight
14.15 lb (5.6 kg)