| AC Test Mode |
| Test frequency |
| PM 6306: |
50, 60, 100, 120 Hz |
| |
200 Hz to 100 kHz (100 Hz steps) |
| |
100 kHz to 1 MHz (1 kHz steps) |
| PM 6304: |
50, 60, 100, 120 Hz |
| |
200 Hz to 20 kHz (100 Hz steps) |
| |
100 kHz |
|
| Test frequency accuracy |
| PM 6306: |
0.01% |
| PM 6304: |
0.01% |
|
| Test signal levels |
| PM 6306: |
50 mV to 2V (10 mV steps) via 100Ω |
| PM 6304: |
50 mV via 100Ω |
| |
1V via 100Ω |
| |
2V via 400Ω |
|
| Basic measurement accuracy at normal measurement mode |
| PM 6306: |
0.1% ± 1 digit (for ³ 0.25V, £ 50 kHz) |
| |
0.1% * (f/50 kHz) ± 1 digit (for ³ 0.25V, > 50 kHz) |
| |
0.1% * (0.25V/Vt) ± 1 digit (for < 0.25V, £ 50 kHz) |
| PM 6304: |
0.05% ± 1 digit (for PM 6304C, £ 2 kHz) |
| |
0.1% ± 1 digit (for £ 20 kHz) |
| |
0.4% ± 1 digit (100 kHz) |
| |
0.5% ± 1 digit (for 50 mV £ 20 kHz) |
| |
2.0% ± 1 digit (for 50 mV, 100 kHz) |
|
| DC bias |
| Internal |
| PM 6306: |
0 to 10V (0.1V steps) |
| PM 6304: |
2V |
|
| External |
| PM 6306: |
0 to 40V |
| PM 6304: |
0 to 40V |
|
| DC test mode (Optional) |
| Test signal levels |
| PM 6306: |
50 mV to 2V (10 mV steps) via 100Ω |
| PM 6304: |
300 mV via 100Ω |
| |
1V via 100Ω |
| |
2V via 400Ω |
|
| Basic measurement accuracy |
| PM 6306: |
0.1% ± 1 digit (for ³ 0.25V) |
| PM 6304: |
0.1% ± 1 digit (for 1V) |
|
| Contact check (PM 6306 only) |
| Pass |
|
| Fail |
| |
3Ω (with indication of failed connection lead) |
|
| Maximum measuring ranges |
| Impedance/Resistance AC Z or RAC |
|
| Resistance DC R DC |
|
| Capacitance C |
|
| Inductance L |
|
| Quality factor Q |
|
| Dissipation factor D |
|
| Phase Angle F |
|
| Voltage monitor VX |
|
| Current monitor IX |
|
| Maximum resolution |
| Impedance/Resistance AC Z or RAC |
|
| Resistance DC R DC |
|
| Capacitance C |
|
| Inductance L |
|
| Quality factor Q |
|
| Dissipation factor D |
|
| Phase Angle F |
|
| Voltage monitor VX |
|
| Current monitor IX |
|
| Circuit diagram |
| Display |
| |
1 of 7 different equivalent circuit diagrams |
|
| Auto Mode |
| Read-out |
| |
Dominant and secondary parameter |
|
| Equivalent circuit diagram |
| |
Parallel for R+C, Serial for R+L |
|
| Manual mode |
| Read-out |
| |
Dominant and secondary parameter or Z, F, D, Q, VX, IX |
|
| Equivalent circuit diagram |
| |
Parallel or serial selectable |
|
| Average function |
| Function |
| |
Exponential averaging in continuous mode |
|
| Levels |
| PM 6306: |
3 (and off) |
| PM 6304: |
1 (and off) |
|
| Deviation mode (PM 6306 only) |
| Relative range in respect to reference value |
|
| Measuring modes (Normal) |
| Continuous |
|
| Single |
| |
Triggered via "TRIG" key |
| |
Triggered via handler interface |
| |
Triggered via IEEE-488 or RS-232 |
|
| Test frequency |
| PM 6306: |
50, 60, 100, 120 Hz |
| |
200 Hz to 100 kHz (100 Hz steps) |
| |
100 kHz to 1 MHz (1 kHz steps) |
| |
DC (optional) |
| PM 6304: |
50, 60, 100, 120 Hz |
| |
200 Hz to 20 kHz (100 Hz steps) |
| |
100 kHz |
| |
DC (optional) |
|
| Read-out |
| |
Display or via IEEE-488 or RS-232 interface |
|
| Measuring Modes (Fast) |
| Max. speed |
|
| Test frequency |
| PM 6306: |
200 Hz to 100 kHz (200 Hz steps) |
| |
100 kHz to 1 MHz (1 kHz steps) |
| |
DC (optional) |
| PM 6304: |
200 Hz to 20 kHz (200 Hz steps) 100 kHz |
| |
DC (optional) |
|
| Single |
| |
Triggered via handler interface |
| |
Triggered via IEEE-488 or RS-232 |
|
| Read-out |
| |
Via IEEE or RS-232 interface (display blanked) |
|
| Binning |
| Standard bins |
|
| Special bins |
|
| Bin programming via |
| |
IEEE-488 interface |
| |
RS-232 interface |
| |
Bin programmer (PM 6304 only) |
|
| Bin limit programming |
|
| Trim function |
| Open circuit |
| |
Open circuit compensation Z > 100 kW |
|
| Short circuit |
| |
Short circuit compensation Z < 10W |
|
| Protection against charged capacitors |
| C < 2 mF |
| PM 6306: |
Vmax < 200V |
| PM 6304: |
Vmax < 500V |
|
| 2 mF o C o 2 mF |
| PM 6306: |
Vmax <47 x (C/mF) -0.234 |
| |
(Vmax in V, C in mF) |
| PM 6304: |
Vmax <117 x (C/mF) -0.234 |
| |
(Vmax in V, C in mF) |
|
| Stored settings (non-volatile memory) |
| Front panel settings |
| |
9 + 1 (trim figures included) |
|
| Bin settings |
|
| Print measurements results |
| Print measurements results |
| |
Via RS-232 interface for serial printers |
|
| Calibration interval |
| Calibration interval |
|
| Environmental Conditions |
| Operating Temperature |
|
| Storage Temperature |
|
| Power, battery life |
|
| Line frequency |
|
| Power Consumption |
| PM 6306: |
44 VA |
| PM 6304: |
31 VA |
|
| Safety |
| |
According to CE-regulation 73/23 EN61010-1 CAT II, Pollution Degree 2 |
|
| EMC |
| |
According to CE regulation 89/336: Emission according to EN 55011, Group 1 Class B, respectively CISPR 11. Immunity according to EN 50082-1, inclusive IEC 801-2, -3, -4 |
|
| Warm up Time |
| PM 6306: |
30 minutes |
| PM 6304: |
5 minutes |
|
| Dimensions and weight |
| W x H x D |
| |
315 x 105 x 405 mm (12.4 in x 4.13 in x 15.9 in) |
|
| Weight |
| PM 6306: |
5.3 kg (11.7 lb) |
| PM 6304: |
4.7 kg (10.4 lb) |
|
|
|